1.Mircowave Communication Laboratory |
Antenna Chamber Room : 325 cm (W) x 285 cm (H) x 640 cm (L) , 8 00 MHz to 18 GHz
|
1.Quiet Zone Size : > 70 cm @0.9 GHz , > 50 cm @1.8 GHz , > 44 cm @2.4 GHz , > 28 cm @5.8 GHz , > 16 cm @18 GHz.
2.Quiet Zone Ripple : < +/- 0.5 dB @1.5~2.4 GHz , < +/- 0.25 dB @2.4~18GHz.
3.Linear and Circular Polarization Gain and Radiation Pattern Measurement. |
Reference Standard : IEEE149, CTIA OTA |
|
 |
2.High Frequency Laboratory |
 |
|
| Agilent 54750+54751+54754 Time Domain Reflectometry Agilent N 5230A PNA + 8753ES Vector Network Analyzer |
Impedance, Attenuation, Propagation Delay, Cross talk, Skew, Risetime Degradation |
Reference Standard : EIA-364-90, EIA-364-101 , EIA-364-103 , EIA-364-108 |
|
|
| |
Item |
Range |
Equipment |
Method |
| High Freq. Lab. |
1.V.S.W.R
2.Return Loss
3.Reflection Coefficient
4.Impedance |
45MHz ~ 6GHz |
Vector Network
Analyzer |
EIA364-108 |
| 1. Attenuation |
45MHz ~ 6GHz |
Vector Network
Analyzer |
EIA364-101 |
|
3.Green Laboratory |
Induced Couple Plasma ICP-OES |
Chemical analysis and environment protection :
1.Lead and Lead Components
2.Cadmium and Cadmium Components
3.Mercury and Mercury Components
4.Chromium |
Reference Standard : EN71 , EN1122 , EPA3050B , EPA3052 , EPA 3060A |
|
 |
Gas Chromatography GC-MS |
Chemical analysis and environment protection :
1.Polybrominated biphenyls(PBBs)
2.Polybrominated diphenylethers(PBDEs) |
Reference Standard : EPA3540C , EPA3550C |
|
 |
Energy Dispersive X-Ray Fluorescence Spectrometer |
Chemical Analysis :
1.Lead
2.Cadmium
3.Mercury
4.Chromium
5.Bromine |
Reference Standard : IEC 62321 |
|
 |
Ion Chromatography |
Chemical Analysis :
1.Flourine
2.Chlorine
3.Bromine
4.Iodine |
Reference Standard : EN61249-2-21 , EN14582 |
|
 |
| |
Item |
Range |
Equipment |
Method |
Green
Product Center |
1.Hg2.Cr3.As4.
Se5.Sb6.Ba |
Fusing |
Inductive
Coupled
Plasma |
EN71 |
| 1. Cd |
Digestion |
EN1122 |
| 1. Pb |
EPA3050B |
|
 |
|
4.Electrical Laboratory |
R&S FSP Spectrum Analyzer ,
Withstanding Voltage Tester ,
Insulation Resistance Meter ,
Low-OHM Meter ,
Temperature Rising Recorder |
| Signal Spectrum Analysis , Withstand voltage , Insulation Resistance , Contact Resistance
Temperature Rising and Safety |
| Reference Standard : EIA364-107 , EIA364-20 , EIA364-21 , EIA364-23
MIL-STD-1344 3001.1 , MIL-STD-1344 3003.1 , MIL-STD-1344 3004.1 |
|
 |
|
5.Mechanical Laboratory |
 |
Vision Measurement Inspection System
Insertion and Extraction Force System
Coating Thickness Measuring System
Torque Gauge
Pull Test Machine
Bending Test Machine |
| Visual Inspection , Insertion and Extraction force , Coating thickness ,
Cable Pull and Bending |
EIA364-13 , EIA364-38 , EIA364-41 , MIL-STD-1344 2013.1 , MIL-STD-1344 2017 |
|
|
6.Environmental Laboratory |
Temperature And Humidity Chamber
Thermal Shock Test Chamber
Forced Convection Oven
Salt Spray Tester
Steam Aging Tester
IR Reflow
|
| Temperature life and Cycling , Thermal Shock , High Temperature , Slat Spray , Solderabilit y |
Reference Standard : EIA364-31 , EIA364-32 , EIA364-17 , EIA364-26
MIL-STD-1344 1002.2 , MIL-STD-1344 1003.1 , MIL-STD-1344 1005.1
MIL-STD-1344 1001.1 |
|
 |
|
7.Vibration Laboratory |
 |
| Mechanical Vibration System |
| Vibration |
Reference Standard :
EIA364-28 , MIL-STD-1344 2005.1 |
|
8.EMS Laboratory |
 |
 |
 |
| Temp. and Humidity Chamber |
Drop Tester |
Mechanical Vibration System |
9.Optical Laboratory |
 |
 |
 |
| Chroma Meter |
Optical Spectrum Analysis |
Integral Sphere Test System |
10.EMC Laboratory (Set up 2009) |
 |
 |
 |
 |
 |
| Vector Network Analyzer |
Wideband Scope |
Pulse Generator |
Spectrum Analyzer |
EMI Probe Kit |
11.Automotive Electronics Lab |
Mechanical Test
Color Spectrophotometer
Coating Thickness System
Wear tester |
Environmental Test
Temperature and Humidity Chamber
Resistance of rain Chamber
Thermal Shock Test Chamber |
Electrical Test
Insulation Resistance Meter
Low-Ohm Meter
LED light Parameter Measuring Machine |
 |
 |
 |
|
|
|
|
|